Frequently
Asked
Questions...
Is there any routine maintenance that should be
done to System 2000 dilution control units?
Yes,
to keep the e-gap eductors running smoothly, we recommend de-liming
them once a year. The procedure is very simple.
- Turn
off the water supply.
- Remove
chemical inlet tube assembly from hose barb.
- Remove
discharge tube from bottom of eductor.
- Unscrew
and remove eductor from water valve assembly.
- Soak
eductor in a 1:4 solution of D-Limer
(#294) for 5 minutes.
- Rinse
off eductor with water.
- Screw
eductor back onto unit.
- Reattach
4 GPM discharge hose, hose barb and chemical inlet tube assembly to
eductor.
- Turn
on water and test unit.
|